About this model

Automated flash purification system with two column channels, four solvent inlets, 1 to 300 mL/min at up to 30 bar, UV (200-400 nm) or UV-Vis (198-810 nm) detection and RFID rack handling. Optional Spektra software licence.

This unit

Biotage Selekt automated flash purification system, manufactured 2022, with two column channels, four solvent inlets, flow from 1 to 300 mL/min at up to 30 bar, and RFID rack handling. Condition and testing: Used, returned from lease or rental. Sold without warranty. Software and computer: Runs from the on-board touchscreen software, so no PC is needed.

Equipment History & Condition

Lease and rental return from a company in Japan or Korea. Ships from South Korea.

What's Included

Biotage Selekt flash purification system

Pre-owned equipment sold by a third-party reseller through MerkaLoop. Ships directly from the reseller.

Specifications

Condition
Good
Year Manufactured
2022
Manufacturer
Biotage
Model
Selekt
Max Pressure
0–30 bar. bar
Ports
Ethernet port, 4 USB ports, 2 D-Sub 9-pin ports, and AUX port.
Power
Electrical supply: 100–127, 220–240 VAC, 50/60 Hz. Connect only to a grounded outlet. Fuses: T4A-250 VAC, P/N 411916. Max power consumed: 300 VA.
Spektra
Add-on software licence for the Selekt family. Lambda-all scans all wavelengths at once for full spectral analysis of peaks as they elute; lambda-all and two individual wavelengths can be shown at once; post-run processing, automatic baseline correction and wavelength focusing.
Firmware
Pump V0.13.25, Controller V1.18.64, UV detector V0.10.22, RFID V1.06.04
Interface
15.0"solvent stable TFT LCD touch screen.
Data Entry
Touchscreen or USB connection to mouse/keyboard.
Rack Types
13 x 100 mm, 16 x 100 mm, 16 x 150 mm, 18 x 150 mm, 25 x 150 mm, 120 mL, 240 mL, and 480 mL.
Part Numbers
SEL-2SV = Biotage Selekt, Two Channel, Single Collection Bed, UV Detector; SEL-2SW = Biotage Selekt, Two Channel, Single Collection Bed, UV-VIS Detector
Certification
CE and FCC marked. CB and NRTL certified.
Network Access
Included
Solvent Supply
A maximum of 4 x 5 liter reservoirs on the optional secondary solvent containment. Larger reservoirs than 5 liters must be placed elsewhere. Note: To be compliant with the secondary containment regulations, do not use containers larger than 4 liters.
Column Channels
Two
Flow Rate Range
1–300 mL/min, in 1 mL/min increments.
Solvent Channels
Four solvent lines; binary gradient with an additional modifier
Wetted Materials
stainless steel, FFKM, PEEK, PP, PE, FEP, PTFE, silica, carbon fiber filled PTFE, and fused silica
Column Size Range
Up to 340 g; up to 1.85 kg with an added safety valve
Gradient Precision
±2%.
Rack Identification
Rack Handling + RFID
Instrument Weight Kg
23–25 kgdepending on the system configuration
Operating Conditions
Location: A laboratory fume hood with the capacity to handle leakage of solvents and provide suitable airflow to prevent the build-up of flammable vapors. Ambient temperature: Operating: 15–32 °C. Storage and transportation: -25–60 °C. Humidity: 20–95% RH. Altitude: Operation: up to 2000 m above sea level. Pollution degree 2.
Column Identification
QR Code Scanner
Instrument Dimensions
Footprint:335/550 mm x 393 mm. The width depends on using one or two collection trays. Height: 545 mmexcluding secondary solvent containment.
Software Compatibility
On-board UNIX-based Java interface with full on-screen QWERTY keyboard. Optional: Biotage Spektra Software Package, Selekt Remote Control, Selekt Monitor API.
Max Number Of Fractions
144 fractions with no rack changeusing 13 x 100 mm racks.
Max Total Fraction Volume
3,840 mL with no rack change, 480 mL bottle racks.
Internal Detector Wavelength
200–400 nmor 198–810 nm. Amplitude range: 0 to 6400 mAU; accuracy of ±5 mAU.
Technique
Flash

Shipping & Logistics

MerkaLoop coordinates all logistics. Timelines depend on seller location and your destination.

  • MerkaLoop coordinates all logistics
  • Shipping quoted before payment
  • Ships globally

Shipping from: South Korea

It is your responsibility to check local regulations and requirements for the handling, acceptance, and servicing of secondhand and new equipment.

contact MerkaLoop about this equipment